MDDT proposal draft

by Qi Gong, Brandon D. Gallas

Product Description

eeDAP is being proposed as an MDDT for Clinical Outcome Assessments (COA). eeDAP is an evaluation environment for digital and analog pathology. eeDAP is a software and hardware platform for designing and executing digital and analog (microscope) pathology studies where the digital image is registered to the real-time view of the corresponding glass slide on the microscope. This registration allows for different pathologists to evaluate the same fields of view (FOVs) in digital mode or in microscope mode. Consequently, it is possible to reduce or eliminate a large source of variability in comparing these modalities in the hands of the pathologist: the FOVs (the tissue) being evaluated. In fact, the current registration precision of eeDAP allows for the evaluation of the same individual cell in both domains. As such, a study can be designed where pathologists are asked to evaluate a preselected list of individual cells or specific FOVs in the digital mode and with the microscope. Consequently, paired observations from coregistered FOVs are collected allowing for a tight comparison between WSI and optical microscopy.

A reader study with eeDAP is intended to evaluate the scanned image, not the clinical workflow of a pathologist or lab. Instead of recording a typical pathology report, eeDAP enables the collection of explicit evaluation responses (formatted data) from the pathologist corresponding to very narrow tasks. This approach removes the ambiguity related to the range of language and the scope that different pathologists use in their reports.

Reader studies utilizing eeDAP are meant to focus on tasks related to specific histopathology features. Since certain image features can challenge image quality properties (color fidelity, focus quality, and depth of field), reader studies with tasks based on features can provide valuable information for the assessment of WSI and its role in clinical practice. eeDAP allows for the formulation of different types of tasks, many of which are currently available in eeDAP: free-text, integer input for counting tasks, a slider in a predefined range for a confidence scoring task (ROC task, receiver operating characteristic task), check boxes of specific categories for a classification task, and marking the image for a search task. These simple tasks can be customized with moderate MATLAB programming skills to suit the study’s purposes.

In what follows, we provide a quick summary of the eeDAP hardware and software. More details can be found in Reference 1.

The eeDAP hardware includes a microscope with a camera (mounted for simultaneous viewing with the eyepiece), a motorized stage (programmable with a stage controller and joystick), and computer as shown in Figure 1. There is also a reticle in the microscope eyepiece. The reticle is needed during image registration, and it is also used to help reduce the area of the FOV or to specify individual cells to evaluate. eeDAP can support usb and FireWire (IEEE 1394) cameras. The camera pixel sizes tested are smaller than 10um. At 40x magnification, this corresponds to camera resolutions better than 0.25 um/pixel. Regarding the motorized stage, eeDAP supports communications with Ludl and Prior stages. The step size for the stages tested are smaller than 0.1 um.

hardware_label.png


The eeDAP software is made up of graphical user interfaces (GUIs) written in MATLAB (Mathworks, Natick, Massachusetts). Using eeDAP does not require a full licensed version of MATLAB. It can be run as a precompiled stand-alone application. The precompiled stand-alone application requires that the free MATLAB compiler runtime (MCR) library be installed (Reference 2). The software uses the Bio-formats library to read WSI images and extract FOVs. Normally the resolution of 40x WSI images is about 0.25 um/pixel. Bio-formats supports several proprietary WSI image formats such as svs, ndpi and tiff. Previous versions of eeDAP used ImageScope, a product of Aperio (a Leica Biosystems Division), which contains an ActiveX control named TIFFcomp to read proprietary WSI images. The change to bio-formats was made because it is better supported, it has fewer limitations (TIFFcomp could only be used with 32-bit MATLAB on Windows), and it is available under the GNU public copyleft licenses.

The eeDAP software is made up of three GUIs: study initialization, global image registration, and data collection. During study initialization, eeDAP reads in an input file. The input file contains the filenames of the WSIs, hardware specifications, and the list of tasks with corresponding FOV locations that will be interpreted by the pathologist. At the end of the study initialization, eeDAP extracts all the WSI FOVs for fast access and transforms the colors so that the image viewed in eeDAP is the same as the image viewed on the scanner-specific viewer.

Global image registration is only done if eeDAP is run in MicroRT mode. The global image registration is equivalent to finding the mathematical relationship between the stage coordinates and WSI coordinates. The global image registration requires three anchors, three pairs of stage-WSI registered coordinates. Each anchor is generated by a local registration: an (x,y) stage coordinate and an (x,y) WSI coordinate that correspond to the same specimen location. The local registration is an interactive process with a study administrator. The study administrator navigates the microscope to a landmark, takes a snapshot of the microscope FOV, and eeDAP records the stage coordinates. Then the study administrator clicks on the corresponding location in the WSI image displayed in the GUI. eeDAP then finds the local registration by maximizing the match of the camera image and WSI.

Full-Figure4_eeDAPregistration.png

The data collection GUI is the same for digital and microscope modes. The GUI shows the WSI FOV and has interfaces for collecting the pathologist’s responses. The difference between modes is that, when run in digital mode, the pathologist sits at the computer and interacts with the data collection GUI. In microscope mode, a study administrator sits at the computer and interacts with the data collection GUI. Meanwhile, the pathologist is engaged with the microscope. The pathologist speaks his or her responses for the study administrator to enter into the GUI. The study administrator also has the responsibility of verifying that the microscope is accurately registered. If it is not, there are buttons to perform an automated registration.

Figure 2.

For the cameras and stages tested, we are able to repeatably and reliably register the WSI image and the glass slide so that the pathologists can evaluate the same FOVs in both modes.


software_label.png



=Context of Use= 1) The device or product area for which the MDDT is qualified
Digital pathology (DP) incorporates the acquisition,
management, and interpretation of pathology information generated from a digitized glass slide. DP is enabled by technological advances in whole slide imaging (WSI) systems, also known as virtual microscopy systems, which can digitize whole slides at microscopic resolution in a short period of time1. The main objective of WSI system is to build digital slides with high magnifications2. The imaging chain of a WSI system consists of multiple components including the light source, optics, motorized stage, and a sensor for image acquisition. WSI systems also have embedded software for identifying tissue on the slide, auto-focusing, selecting and combining different fields of view (FOVs) in a composite image, and image processing (color management, image compression, etc.). Details regarding the components of WSI systems can be found in Gu and Ogilvie3.
__Q for WSI group: We hope to get background, introduction and meaning about Digital pathology, WSI and devices.__

2) The stage(s) of device development (e.g., early feasibility study, pivotal study, etc.)
In the United States, WSI device is available in
market for researching and education, but it has not been approved for clinical usage. However, it got the clinical approval in some other countries such as Canada, Sweden several years ago. And recently, it got the approval in The UK and Ireland.
__Q for WSI group and manufacture: We hope to get more information about development stage of DP, WSI. For example several big countries’ development stage and when did they approved the stage.__

3) The specific role of the MDDT (for clinical uses this includes the study population or disease characteristics, as well as specific use – diagnosis, patient selection, clinical endpoints).”
The eeDAP is proposed as an MDDT for evaluating WSI performance and collecting labels for training image analysis programs.

4) How to use MDDT in regulatory context
The eeDAP is proposed as an MDDT for the context of use measuring and evaluating quantity of WSI. It is used to collect pathologists’ evaluations for the same FOV’s on WSI (Digital Mode) and glass slide (MicroRT Mode), at the same time asks expert to do some tasks about the ROIs. The measurement and evaluation of quantity of WSI are based on the similarity our two modes results.



Strength of Evidence

  • Tool Validity. Does the available data adequately support the validity

of the measurement? Does the MDDT measure reliably and accurately? Depending on the tool type, this may include analytical, clinical, and construct validity, sensitivity, specificity, accuracy, precision, repeatability, external validity, reduction of bias, verification of the constitutive model, uncertainty quantification, numerical convergence, etc.

  • Plausibility. Is it scientifically plausible that the measurements obtained through use of the MDDT are related to the true outcome of interest? Is there a causal path or mechanistic explanation to connect the MDDT to the outcome?
  • Extent of Prediction. What data are available to demonstrate a predictive relationship between the MDDT and the true outcome of interest? What is the strength of that predictive relationship? Is the prediction repeatedly demonstrated in multiple studies or as a class effect? If relevant, is the conclusion (that the effect of treatment on the measurement obtained using the MDDT predicts the outcome of interest) supported by credible information?
  • Capture. Does the MDDT fully capture the aggregate effect of the intervention on the true outcome of interest? Does the MDDT account for every major effect of the intervention? Are there available data which call this into question?

Plan for get data

1. Registration quality (Tool Validity)

Create new task in eeDAP to test it.

  • Question: When does registration succeed?
  • Options: A. Automatically B. After focus and fast registration. C.

After best registration. D. Fail.

  • Measurement: Whether the center of reticle is inside target cell.
2. Color display quality (Tool Validity)
Apply Image ICC and manufacture viewer software ICC to eeDAP (base on Bioformat).
Measure the output signal from cable for eeDAP WSI and manufacture viewer software WSI.

Q for Manufacture: We need viewer software ICC and application method to process our extracted WSI image.

3. Study results (Capture)
Complements NIH mitotic counting study
  • Calculate the reader agreement (bias and correlation).

Q for NIH: We hope to get all mitotic figure counting data (ROIs, results).
Q for WSI group: We hope to find experts from the group to do mitotic figure counting.


Assessment of Advantages and Disadvantages

Advantages:

1. Eliminates location variability
The eeDAP system can help experts read exact same ROIs on WSI and a glass slide.
2. Reading efficiency
After one time registration between WSI and glass slide, all ROIs on this specimen moving is automatic.
3. Results collecting and analyzing
Output report includes all study settings and results in a neat format.
4. Customizable
Users can create new kinds of task if they do Matlab program.

Disadvantages:

1. Not the native viewer
The eeDAP is WSI image is extract and display by Bioformat and Matlab bur not manufacture viewer software. The influence can be reduced by applying ICC.
2. Current not pan and zoom
Currently, user can’t pan and zoom WSI in eeDAP.
3. Lack of global view
Currently, all existing tasks are based on FOV not whole slide.


Consent to Public Disclosure and Use

The compiled eeDAP software and all source codes are shared on Github with HTML and PDF versions manuals. The video version manual is under processing. The only prerequisite software for compiled eeDAP software is MATLAB compiler runtime, which is free to download and use. These guarantee (1) FDA can make public sufficient information to support use of the qualified eeDAP and (2) the general public can use the eeDAP and rely on data generated using the eeDAP in gaining FDA clearance or approval of other devices.


References

  1. B. D. Gallas et al. “Evaluation Environment for Digital and Analog Pathology: A Platform for Validation Studies.” Journal of Medical Imaging 1.3 (2014).
  2. https://www.mathworks.com/products/compiler/mcr.html
  3. https://www.openmicroscopy.org/site/support/bio-formats5.3/
  4. M. G. Rojo et la. “Critical Comparison of 31 Commercially Available Digital Slide Systems in Pathology.” International Journal of Surgical Pathology, 14(4), 285-305. (2006).
  5. J. Gu and R. W. Ogilvie, Eds., Virtual Microscopy and Virtual Slides in Teaching, Diagnosis, and Research, CRC Press, Boca Raton, Florida (2005).